Back to Top

Hi, Guest!

  LOKASI :  Kabupaten Banyumas

Bergabung Selama :

BAGIKAN :   

Bagikan :

SYD-3536D Automatic Cleveland Open Cup Flash Point Tester

Update Terakhir
:
01 / 12 / 2019
Min. Pembelian
:
0
Dilihat Sebanyak
:
9 kali

Harga

CALL
Bagikan
:
Kami adalah perusahaan Distributor yang bergerak di bidang penjualan berbagai macam alat alat pertanian dan kesehatan dan berbagai macam alat ukur kadar air, termometer infra merah, dan alat ukur alat uji lainnya

Perhatian !

Perusahaan ini terdaftar sebagai Free Member. Hindari melakukan pembayaran sebelum bertemu penjual atau melihat barang secara langsung. COD (Cash On Delivery) atau bertemu langsung dengan penjual merupakan metode transaksi aman yang kami sarankan.

Detail SYD-3536D Automatic Cleveland Open Cup Flash Point Tester

Summary This instrument is designed and made as per GB/ T3536-2008 Petroleum Products - Determination of Flash and Fire Points - Cleveland open cup method and ASTM D 92 Standard Test Method for Flash and Fire Points by Cleveland Open Cup Tester. It is applicable to all petroleum products with flash points above 79 and below 400 except fuel oils. I. Main technical features 1. This instrument is a kind of fully-automatic instrument which can do determination for both flash point and fire point. 2. This instrument adopts technology of single chip machine and LCD screen. Full English man-machine dialog interface. It can preset the parameters of expected flash point, sample mark number, atmospheric pressure, test date, etc. It has menu to prompt and input function of guide type. 3. This instrument adopts desktop structure and advanced design. The man-machine interface make it easy to use. And the test result is accurate. It can be widely used in railway, aviation, electric power, petroleum industry and colleges, scientific research institutes, measurement departments to do determination of flash point and fire point for petroleum products. II. Main technical specifications and parameters 1. Power supply: AC ( 220 10% ) V, 50Hz 2. Temperature measurement: Range: room temperature~ 400 ; Repeatability: d8 ; Reproducibility: d17 ; Accuracy: 0.1 3. Heating rate: Correspond with GB/ T3536 and ASTM D92 4. Igniting mode: Electric ignition, Gas flame diameter: 3.2mm~ 4.8mm 5. Ambient temperature: ( 10~ 40)  6. Relative humidity: d80% 7. Total power consumption: d500W 8. Size: 520mm 380mm 310mm; Weight: 13kg
Tampilkan Lebih Banyak